These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Safe use of heparin-coated bypass circuits incorporating a pump-oxygenator.
    Author: Jones DR, Hill RC, Vasilakis A, Hollingsed MJ, Graeber GM, Gustafson RA, Cruzzavala JL, Murray GF.
    Journal: Ann Thorac Surg; 1994 Apr; 57(4):815-8; discussion 818-9. PubMed ID: 8166524.
    Abstract:
    Durable, covalently bonded, heparin-coated cardiopulmonary bypass (CPB) circuits with oxygenators have been developed. Proposed advantages of heparin-coated CPB circuits include improved biocompatibility and thromboresistance. The purpose of this study was to evaluate our experience with heparin-coated CPB circuits in 20 patients. Heparin was given to maintain an activated clotting time equal to or greater than 200 seconds, while flow rates were kept equal to or greater than 2 L/min. Indications for use of this circuit included recent stroke, posttraumatic injuries, recent gastrointestinal bleeding, protamine allergies, combined cardiac and noncardiac procedures, and ventricular assist. Mean heparin dosage was 0.50 +/- 0.18 mg/kg and protamine dosage was 57.14 +/- 39.36 mg. Postoperative blood loss and transfusion requirements were minimal. Postoperative complement levels of C3a and C5a were normal, suggesting excellent biocompatibility. There were no deaths or perioperative complications. Heparin-coated CPB circuits using a pump oxygenator can be used safely with low-dose heparin administration in select patients requiring CPB.
    [Abstract] [Full Text] [Related] [New Search]