These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Balloon expandable intravascular stents: aortic implantation and late further dilation in growing minipigs.
    Author: Grifka RG, Vick GW, O'Laughlin MP, Myers TJ, Morrow WR, Nihill MR, Kearney DL, Mullins CE.
    Journal: Am Heart J; 1993 Oct; 126(4):979-84. PubMed ID: 8213458.
    Abstract:
    Intravascular stents have been implanted in children with congenital and acquired vascular stenoses. Little information is known regarding the long-term results in growing patients, the implantation of multiple stents (in series), and the feasibility of further dilation. Nine stents were implanted in the abdominal aortas of 6 mini-pigs; in 3 pigs, 2 stents were implanted in series. Five pigs were recatheterized 196 +/- 17 days after stent implant. All stents were patent, with mild narrowing at the stent relative to the adjacent aorta; further stent dilation was performed. The stent diameter increased from 8.3 +/- 0.6 mm to 11.2 +/- 1.3 mm (p = 0.002), and the narrowing was relieved. These five pigs were catheterized 253 +/- 73 days after stent redilation. Each stent was patent without narrowing or pressure gradient. We conclude that intravascular stents do not interfere with normal growth, can be implanted in series, and can be further dilated.
    [Abstract] [Full Text] [Related] [New Search]