These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: The retinal irradiance and spectral properties of the multiport illumination system for vitreous surgery. Author: Koch FH, Schmidt HP, Mönks T, Blumenröder SH, Haller A, Steinmetz RL. Journal: Am J Ophthalmol; 1993 Oct 15; 116(4):489-96. PubMed ID: 8213981. Abstract: The multiport illumination system was developed as a fiberoptically illuminated pars plana cannula system for vitreous surgery. The retinal irradiance of the multiport illumination system and several other commonly used endoilluminators (light pipes) measured at 5 mm are comparable, ranging from 94 to 256 mW/cm2. In the actual surgical setting, however, the retinal irradiance of the multiport illumination system is far less (56 mW/cm2) because the end of the fiberoptic cables are fixated in the pars plana 16.5 mm away from the retinal surface. A reduction in the maximal intensity (100%) to 50% reduces the retinal irradiance by approximately two thirds to 3.7 mW/cm2. The light emitted by the multiport illumination system used in conjunction with a long-wave pass filter with a cutoff wavelength of 420 nm reduces the amount of short-wavelength light by one third to about 4% of the overall irradiance.[Abstract] [Full Text] [Related] [New Search]