These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Retrieval of a migrated coronary stent by means of an alligator forceps catheter. Author: Eeckhout E, Stauffer JC, Goy JJ. Journal: Cathet Cardiovasc Diagn; 1993 Oct; 30(2):166-8. PubMed ID: 8221873. Abstract: Stent migration and embolization are well-known complications of intracoronary stenting with balloon-mounted stents. During an elective stenting procedure of a proximal right coronary artery stenosis, a 3.5 mm Wiktor stent (Medtronic Inc., Minneapolis) was displaced from its delivery balloon. The guiding catheter and the delivery system were withdrawn, leaving the stent around a 3 m 0.014 inch High Torque floppy guidewire (ACS, Santa Clara, CA) in the abdominal aorta. An 40 cm 5F Alligator Forceps catheter (Cook OB/Gyn., Spencer, IN), introduced through a cut-off 8F coronary guiding catheter, allowed improved torque control of the retrieval catheter and a safe and successful withdrawal of the stent through the arterial introducer sheet.[Abstract] [Full Text] [Related] [New Search]