These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Quantitative analysis of four EMG amplifiers.
    Author: Perreault EJ, Hunter IW, Kearney RE.
    Journal: J Biomed Eng; 1993 Sep; 15(5):413-9. PubMed ID: 8231159.
    Abstract:
    Four typical EMG amplifiers were tested quantitatively to observe the diversity and specificity of available equipment. Gain, phase, common mode rejection ratio (CMRR) and noise characteristics were measured for each device. Various gain and phase responses were observed, each best suited to specific application areas. For all amplifiers, the CMRR was shown to decrease dramatically in the presence of input impedance mismatches of more than 10 k omega between the two electrodes. Because such impedance mismatches are common on the skin surface, these results indicate that proper skin preparation is required to maximize the noise rejection capabilities of the tested amplifiers.
    [Abstract] [Full Text] [Related] [New Search]