Title: In vivo long-term evaluation of the Utah electrohydraulic total artificial heart. Author: Tatsumi E, Khanwilkar PS, Rowles JR, Chiang BY, Burns GL, Long JW, Hansen AC, Holfert JW, Bearnson GB, Crump KR. Journal: ASAIO J; 1993; 39(3):M373-80. PubMed ID: 8268562. Abstract: An electrohydraulic total artificial heart (EHTAH) has been developed and evaluated by long-term in vivo studies. The EHTAH is composed of blood pumps with an interatrial shunt (IAS), an energy converter, and electronics. The EHTAH with external electronics was implanted in four calves weighing from 81-90 kg. Two animals died on the 1st and 5th post operative days, the third animal survived for 32 days, and the fourth for 159 days. The IAS was free of thrombus at autopsy in all animals. The longest surviving animal increased in size from a pre operative weight of 81 kg to 134 kg on day 144. Cardiac output ranged from 9.3 to 10.5 L/min, whereas right and left atrial pressures increased with the calf's growth from 4-10 to 16-20 mmHg and from 8-14 to 18-22 mmHg, respectively. The animal favorably tolerated up to 3.4 km/hr of treadmill exercise, both hemodynamically and metabolically. The elevation of atrial pressures during treadmill exercise was significantly alleviated by employing an automatic control mode. It is concluded that the device has the potential to be a totally implantable system for permanent use.[Abstract] [Full Text] [Related] [New Search]