Title: Comparison of image quality and mean absorbed dose to the breast for two mammographic films. Author: Persliden J, Fransson V, Vitak B, Fagerberg G. Journal: Acta Radiol; 1993 Jul; 34(4):351-5. PubMed ID: 8318296. Abstract: A film intended for mammography (Kodak SO 155 MRH-1) was tested and compared to one (Kodak SO 177 Ortho M) used earlier. Both films including the cassette could resolve 20 lp/mm. For a processing time of 90 s the new film gave the same overall image quality and irradiation dose to the breast as the old system. If, instead, the processing time was increased to 150 s, a 43% reduction in kerma could be attained with the new film. With a developing temperature of 36 degrees C and 150 s processing time, the noise is clinically acceptable. A nonparametric test showed no significant difference between the 2 films on the 0.01 level. At an X-ray tube potential difference of 25 kV, the mean absorbed dose to a 4.5-cm-thick breast was reduced from 1.7 mGy with the old combination to 1.0 mGy with the new one. The measurements were made with a moving grid.[Abstract] [Full Text] [Related] [New Search]