These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: High-performance liquid chromatographic determination of amoxicillin in urine using solid-phase, ion-pair extraction and ultraviolet detection.
    Author: Chulavatnatol S, Charles BG.
    Journal: J Chromatogr; 1993 May 19; 615(1):91-6. PubMed ID: 8340467.
    Abstract:
    A simple, robust high-performance liquid chromatographic method is described for assaying amoxicillin in urine. Sample clean-up involved solid-phase, ion-pair extraction onto Sep-Pak C18 cartridges followed by elution with acetonitrile (3%, v/v) in pH 4.85 buffer. Separations were performed on an Ultrasphere C18 column with a mobile phase comprising acetonitrile (32.5 ml) and pH 7.1 phosphate buffer (0.01 M, 1000 ml). Peaks were detected at 229 nm. Recovery was greater than 94%. Plots of peak area against urinary amoxicillin concentration were linear (r > 0.999) from 5 to 500 mg/l. Between-day and within-day imprecision (coefficient of variation) ranged between 3.0 and 15.5%. Inaccuracy was 7.1%, or less.
    [Abstract] [Full Text] [Related] [New Search]