These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Measurement of ascorbic acid in human aqueous humour and plasma and bovine aqueous humour by high-performance liquid chromatography with electrochemical detection. Author: Watson DG, Iqbal Z, Midgley JM, Pryce-Jones H, Morrison L, Dutton GN, Karditsas S, Wilson W. Journal: J Pharm Biomed Anal; 1993; 11(4-5):389-92. PubMed ID: 8357877. Abstract: High-performance liquid chromatography (HPLC) on ODS silica with cetyltrimethylammonium bromide as an ion-pairing reagent and electrochemical detection (ECD) was used to determine ascorbic acid in bovine and human aqueous humour and human plasma. Hydroquinone was used as the internal standard. A calibration curve plotted with ascorbic acid concentrations in the range 0.5-5 micrograms ml-1 for peak height versus internal standard peak height had a correlation coefficient of 0.998. The RSD (precision) between analyses of the same diluted sample was 1.5% and the RSD (reproducibility) between analyses of separate aliquots of the same sample of aqueous humour was 1.6%.[Abstract] [Full Text] [Related] [New Search]