These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Evaluation of pressure changes in feeding arteries during embolization of intracerebral arteriovenous malformations.
    Author: Handa T, Negoro M, Miyachi S, Sugita K.
    Journal: J Neurosurg; 1993 Sep; 79(3):383-9. PubMed ID: 8360735.
    Abstract:
    The pressure in 47 arteries feeding 21 arteriovenous malformations (AVM's) was investigated during transarterial embolization using a Tracker-18 microcatheter. On average, systolic pressure increased by 22 mm Hg. In AVM's with single or few feeders, embolization was usually achieved well; in contrast, giant AVM's with multiple feeders and a large arteriovenous shunt were poorly embolized. However, large AVM's with well-demarcated components may be reduced by embolization to an appropriate size for surgery or stereotactic radiation therapy. It was found that the feeding artery pressure increased significantly more in well-embolized than in poorly embolized cases. Measurements of the feeding artery pressure clarified the hemodynamics of AVM's and facilitated more successful embolization.
    [Abstract] [Full Text] [Related] [New Search]