These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Automated sample clean-up with solid-phase extraction for the determination of aflatoxins in urine by liquid chromatography. Author: Kussak A, Andersson B, Andersson K. Journal: J Chromatogr; 1993 Jul 02; 616(2):235-41. PubMed ID: 8376505. Abstract: An automated extraction and clean-up procedure was developed for the determination of aflatoxins in human urine at the 50 pg/ml level. Aflatoxins B1, B2, G1 and G2 are captured on C2 extraction columns and simultaneously cleaned up with the aid of a robotic system. The processed samples are analysed by reversed-phase high-performance liquid chromatography. Fluorescence detection was enhanced for aflatoxins B1 and G1 using factorial design optimization of the post-column reactor. Silylation of the glass vials used in the robotic system was of the utmost importance. With non-silylated glass vials, up to 75% of the analytes were lost. Average aflatoxin recoveries were B1 95%, B2 90%, G1 93% and G2 89%.[Abstract] [Full Text] [Related] [New Search]