  • Title: Improved prediction of extracellular and total body water using impedance loci generated by multiple frequency bioelectrical impedance analysis.
    Author: Cornish BH, Thomas BJ, Ward LC.
    Journal: Phys Med Biol; 1993 Mar; 38(3):337-46. PubMed ID: 8451277.
    Abstract:
    Bioelectrical impedance analysis (BIA) using a frequency of 50 kHz is an established method of predicting total body water (TBW). However, very little research has been performed to determine whether 50 kHz is the optimum frequency for the prediction of TBW from impedance measurements. This paper analyses a mathematical expression describing the equivalent electrical circuit for biological tissue, and derives a graphical representation of the resistive and reactive components. The nature of the resulting impedance locus was used in the analysis of measured whole-body impedance of 42 rats over a range of frequencies to determine the impedance at the characteristic frequency, Zc, and also the impedance at zero frequency, R0. The standard error associated with the prediction of TBW (determined by isotope dilution) using Zc was 5.9% compared with a standard error of prediction of 10.1% using the established BIA method at 50 kHz on the same data. Predictions of extracellular water (ECW) using the impedance at zero frequency, R0, yielded a standard error of 3.2% compared with standard errors of 4.8% and 4.2% using single frequency BIA measurements at 5 kHz and 1 kHz, respectively. These results demonstrate a significant (P < 0.01) improvement in the prediction of TBW and ECW using multiple frequency bioelectrical impedance analysis (MFBIA).
    [Abstract] [Full Text] [Related] [New Search]