These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Reproducibility of the signal-averaged, high-pass filtered electrocardiogram].
    Author: Seifert T, Karbenn U, Borggrefe M, Breithardt G.
    Journal: Z Kardiol; 1993 Mar; 82(3):143-9. PubMed ID: 8475650.
    Abstract:
    In order to test the reproducibility of the signal-averaged electrocardiogram (SAECG) using Simson's method (high-pass filter cutoff frequency 25 Hz, orthogonal leads, recording of 133 s per lead), 121 patients were examined. In all patients, two signal-averaged ECGs were performed on the first day of the study immediately after each other, using identical electrode position. In a subgroup of 47 patients, the same procedure was repeated 3 days later. There was no difference between the mean values of conventionally calculated averaging parameters (heart rate, QRS-duration, root-mean-square voltage in the terminal 40 ms of the highly amplified and filtered QRS-complex [V40], the low amplitude signal duration under 40 microV in the terminal portion of the QRS-complex [LAS], total root-mean-square voltage of the QRS-complex), both with regard to immediate and short-term reproducibility. Thus, conventionally calculated averaging parameters are well reproducible.
    [Abstract] [Full Text] [Related] [New Search]