These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Comparison of invasive blood pressure measurement in the aorta with indirect oscillometric blood pressure measurement at the wrist and forearm]. Author: Saul F, Aristidou Y, Klaus D, Wiemeyer A, Lösse B. Journal: Z Kardiol; 1995 Sep; 84(9):675-85. PubMed ID: 8525669. Abstract: Indirectly measured blood pressure at the wrist or upper arm was compared with directly measured values in the aortic arch during routinely performed diagnostic cardiac catheterization in 100 patients (31-80 years, mean 59.3 years, 60% males). The noninvasive measurements were carried out by oscillometric devices, NAiS Blood Pressure Watch for measurements at the wrist, and Hestia OZ80 at the upper arm. Systolic blood pressure measured at the wrist was 4.3 +/- 14.1 mm Hg, and the diastolic value 6.0 +/- 8.9 mm Hg higher than when measured at the aortic arch; the difference was significant in both cases. Correlation coefficients were 0.85 for systolic and 0.71 for diastolic blood pressure. In 16% of the patients the systolic blood pressure at the wrist differed more than +/- 20 mm Hg. The diastolic blood pressure at the wrist measured more than +/- 20 mm Hg higher than in the aorta in 5% of the patients. At the upper arm mean systolic values were not different to the aorta. The diastolic pressure was 9.3 +/- 9.8 mm Hg higher in the aorta than at the upper arm. To verify the accuracy of values measured with the NAiS Blood Pressure Watch compared with the standard technique at the upper arm, sequential measurements were made at wrist and ipsilateral upper arm in the same group of 100 patients. The systolic blood pressure at the left wrist was 3.4 +/- 13.3 mm Hg higher and the diastolic pressure 3.8 +/- 9.5 mm Hg lower than at the upper arm. Only 53% of systolic values lay within a range of +/- 10 mm Hg. The correspondence between wrist and upper arm values was better for diastolic blood pressure, the values differing by less than +/- 10 mm Hg in two-thirds of patients. Self-measurement of arterial blood pressure with an oscillometric device at the wrist can be recommended only in individual cases with a difference of simultaneously measured values at the upper arm of less than +/- 10 mm Hg for systolic and diastolic blood pressures. The standard method for indirectly measuring arterial blood pressure remains the measurement at the upper arm site, which nevertheless showed a systolic pseudohypertension (deviation of more than 10 mm Hg) in comparison to the invasively measured values in 15% of our selected patients and a diastolic pseudohypertension (deviation of more than 15 mm Hg) in 23% of the patients.[Abstract] [Full Text] [Related] [New Search]