These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Detection of malignancies with SPECT versus PET, with 2-[fluorine-18]fluoro-2-deoxy-D-glucose.
    Author: Martin WH, Delbeke D, Patton JA, Sandler MP.
    Journal: Radiology; 1996 Jan; 198(1):225-31. PubMed ID: 8539384.
    Abstract:
    PURPOSE: To compare single photon emission computed tomography (SPECT) with 2-[fluorine-18]fluoro-2-deoxy-D-glucose (FDG) and positron emission tomography (PET) with FDG to evaluate malignancies. MATERIALS AND METHODS: PET and SPECT, with fluorine-18 sodium fluoride, were performed sequentially in a cylindric phantom that contained different size spheres with activity ratios of 5:1, 10:1, and 15:1. PET and SPECT were also performed in 24 patients with known or suspected malignancies. RESULTS: Sensitivities of PET and SPECT were 2,238 cpm/microCi (82.8 cpm/MBq) and 129 cpm/microCi (4.8 cpm/MBq), respectively (reconstructed spatial resolution, 7 and 17 mm, respectively [13-cm radius of rotation]). In the phantom studies, lesions of 1.5 and 1.3 cm or more in diameter were detected with a ratio of 5:1 and 10:1, respectively, and an information density of 150 counts per square centimeter. At FDG PET, 46 hypermetabolic lesions consistent with tumor were depicted in patients; at FDG SPECT, 36 (78%) were depicted. Sensitivity of FDG SPECT was 92% for detection of malignancies 1.8 cm or more in diameter seen at FDG PET. CONCLUSION: Findings at FDG SPECT can help differentiate benign from malignant lesions.
    [Abstract] [Full Text] [Related] [New Search]