These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Evaluation of clinical assays for measuring high-dose methotrexate in plasma.
    Author: Albertioni F, Rask C, Eksborg S, Poulsen JH, Pettersson B, Beck O, Schroeder H, Peterson C.
    Journal: Clin Chem; 1996 Jan; 42(1):39-44. PubMed ID: 8565230.
    Abstract:
    Four routine assays commonly used for monitoring plasma methotrexate (MTX) during high-dose therapy were validated by HPLC as the comparison method. MTX and its main metabolite, 7-hydroxymethotrexate (7-OHMTX), were analyzed by HPLC with postcolumn derivatization and fluorometric detection. About 200 clinical plasma samples from 13 children with acute lymphoblastic leukemia who received 5-8 g/m2 MTX as 24-h infusions were analyzed. The fraction of measured concentrations of MTX that were within 75-125% of the values obtained by HPLC were 64.5% for enzyme inhibition assay, 56.4% for fluorescence polarization immunoassay with polyclonal antibodies (FPIA1; Abbott), 58.9% for FPIA2 (with monoclonal antibodies; Abbott), and 46.4% for enzyme-multiplied immunoassay (Emit; Syva). All nonchromatographic procedures were subject to interferences from MTX plasma metabolites or endogenous substances. The interference from 7-OHMTX was, however, somewhat less pronounced for FPIA2 (monoclonal) than for FPIA1 (polyclonal).
    [Abstract] [Full Text] [Related] [New Search]