These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [High intensification film-screen systems in thoracic diagnosis. A clinical comparative study].
    Author: Schäfer CB, Sokiranski R, Claussen CD.
    Journal: Aktuelle Radiol; 1995 Nov; 5(6):389-93. PubMed ID: 8580141.
    Abstract:
    OBJECTIVE: The quality of chest images was evaluated for a conventional screen-film system, a new asymmetric screen-film system, and a new uv screen-film system. MATERIALS AND METHODS: 138 chest radiographs (69 p.a., 69 lateral) obtained with three different high intensification screen-film combinations were compared. Film density and film contrast were measured. Three readers graded the image quality according to 16 criteria. RESULTS: The asymmetric film-screen combination Insight HC showed the lowest film contrast, the best exposure range, and a elevated film density in the mediastinal area. The asymmetric screen-film system was ranked by all three observers as being substantially better in image quality. CONCLUSION: Compared to conventional screen-film systems, the new screen-film systems can improve the image quality of chest radiographs. Therefore, the new high intensification screen-film combinations can be used as a low cost alternative to the Amber technique and digital radiography.
    [Abstract] [Full Text] [Related] [New Search]