These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Simultaneous determination of catecholamines, their basic metabolites and serotonin in urine by high-performance liquid chromatography using a mixed-mode column and an eight-channel electrochemical detector. Author: Mashige F, Matsushima Y, Miyata C, Yamada R, Kanazawa H, Sakuma I, Takai N, Shinozuka N, Ohkubo A, Nakahara K. Journal: Biomed Chromatogr; 1995; 9(5):221-5. PubMed ID: 8593422. Abstract: A high-performance liquid chromatographic method for simultaneous determination of free catecholamines, their basic metabolites and serotonin in human urine was developed. The compounds were separated on a precolumn of cation-exchange resin and a mixed-mode (C18/cation-exchange) column and determined by an eight-channel electrochemical detector. By this method, free norepinephrine (NE), epinephrine (E), dopamine (DA), normetanephrine (NMN), metanephrine (MN), 3-methoxy-tyramine (3-MT) and serotonin (5-HT) were determined in 40 min with inter-assay precision of 1.9, 3.2, 1.2, 1.6, 1.6, 5.5 and 3.2%, respectively. Detection limits were between 0.05 (E) and 0.3 (3-MT) pmol. The mean values of analytical recovery for NE, E, DA, NMN, MN, 3-MT, 5-HT were 99, 104, 104, 99, 102, 109 and 99%, respectively.[Abstract] [Full Text] [Related] [New Search]