These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Reducing the levels of formaldehyde exposure in gross anatomy laboratories.
    Author: Coleman R.
    Journal: Anat Rec; 1995 Dec; 243(4):531-3. PubMed ID: 8597301.
    Abstract:
    BACKGROUND: A method is described in which formaldehyde levels are greatly reduced in our gross anatomy laboratory in order to comply with increasingly severe safety and health regulations. METHODS: A novel type of dissection "bed" has been introduced which incorporates an internal motor that causes a downflow of formaldehyde-rich vapors, which are absorbed by a replaceable active carbon filtration system. RESULTS: Use of the new dissection "beds" has resulted in the recirculated air being virtually formaldehyde-free. Formaldehyde vapor levels in our gross anatomy laboratory have been greatly reduced and are typically in the range of 0.03-0.09 ppm. CONCLUSIONS: The new system allows us to comply with safety and health regulations and provide a dissection room with an excellent working environment.
    [Abstract] [Full Text] [Related] [New Search]