  • Title: Three routine methods for measuring high-density lipoprotein cholesterol compared with the Reference Method.
    Author: Harris N, Galpchian V, Rifai N.
    Journal: Clin Chem; 1996 May; 42(5):738-43. PubMed ID: 8653900.
    Abstract:
    We compared the performance of three methods for quantifying high-density lipoprotein cholesterol (HDL-C) with the Reference Method for HDL-C, using samples with a wide range of triglyceride (TG) concentrations (290-18000 mg/L). All three comparison assays-- utilizing a magnetic dextran sulfate precipitating reagent, a direct method, and a standard MgCl2-dextran sulfate reagent--were precise, with a run-to-run CV of less than or equal to 4.1%. However, the systematic error of these assays exceeded the National Cholesterol Education Program (NCEP) performance goal of less than or equal to 10% in half of the concentration ranges tested. Nevertheless, the total error of the assays generally meets the current 22% limit set by the NCEP. Although both the magnetic dextran sulfate precipitation reagent and the direct assay can be performed more rapidly than the MgCl2-dextran sulfate assay, the direct assay involves no sample preparation and requires only 4 microL of sample excluding the dead space. Although precipitation is frequently inadequate with the MgCl2-dextran sulfate reagent at TG concentrations >6000 mg/L, both the magnetic and the direct reagent show no interference from high TG concentrations as great as 18 000 mg/L.