Search MEDLINE/PubMed
Title: Clinical laboratory investigation of the Sanofi ACCESS CK-MB procedure and comparison to electrophoresis and Abbott IMx. Author: Mao GD, Adeli K, Eisenbrey AB, Artiss JD. Journal: Am J Clin Pathol; 1996 Jul; 106(1):112-7. PubMed ID: 8701919. Abstract: This evaluation was undertaken to verify the application protocol for the CK-MB assay on the ACCESS Immunoassay Analyzer (Sanofi Diagnostics Pasteur, Chaska, MN). The results show that the ACCESS CK-MB assay total imprecision was 6.8% to 9.1%. Analytical linearity of the ACCESS CK-MB assay was excellent in the range of < 1-214 micrograms/L. A comparison of the ACCESS CK-MB assay with the IMx (Abbott Laboratories, Abbott Park, IL) method shows good correlation r = 0.990 (n = 108). Linear regression analysis yielded Y = 1.36X-0.3, Sx/y = 7.2. ACCESS CK-MB values also correlated well with CK-MB by electrophoresis with r = 0.968 (n = 132). The linear regression equation for this comparison was Y = 1.08X + 1.4, Sx/y = 14.1. The expected non-myocardial infarction range of CK-MB determined by the ACCESS system was 1.3-9.4 micrograms/L (mean = 4.0, n = 58). The ACCESS CK-MB assay would appear to be rapid, precise and clinically useful.[Abstract] [Full Text] [Related] [New Search]