These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Accuracy of a pulse oximeter during hypoxia]. Author: Tachibana C, Fukada T, Hasegawa R, Satoh K, Furuya Y, Ohe Y. Journal: Masui; 1996 Apr; 45(4):479-82. PubMed ID: 8725606. Abstract: The accuracy of the pulse oximeter was examined in hypoxic patients. We studied 11 cyanotic congenital heart disease patients during surgery, and compared the arterial oxygen saturation determined by both the simultaneous blood gas analysis (CIBA-CORNING 288 BLOOD GAS SYSTEM, SaO2) and by the pulse oximeter (DATEX SATELITE, with finger probe, SpO2). Ninty sets of data on SpO2 and SaO2 were obtained. The bias (SpO2-SaO2) was 1.7 +/- 6.9 (mean +/- SD) %. In cyanotic congenital heart disease patients, SpO2 values were significantly higher than SaO2. Although the reason is unknown, in constantly hypoxic patients, SpO2 values are possibly over-estimated. In particular, pulse oximetry at low levels of saturation (SaO2 below 80%) was not as accurate as at a higher saturation level (SaO2 over 80%). There was a positive correlation between SpO2 and SaO2 (linear regression analysis yields the equation y = 0.68x + 26.0, r = 0.93). In conclusion, the pulse oximeter is useful to monitor oxygen saturation in constantly hypoxic patients, but the values thus obtained should be compared with the values measured directly when hypoxemia is severe.[Abstract] [Full Text] [Related] [New Search]