These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Vacuum sources in obstetrics.
    Author: Wiper DW, Duchon MA, Muise KL.
    Journal: J Reprod Med; 1996 Jun; 41(6):444-6. PubMed ID: 8799923.
    Abstract:
    OBJECTIVE: To characterize the performance of vacuum equipment used in operative vaginal deliveries. STUDY METHOD: We measured wall suction in eight Cleveland area labor-and-delivery units. Additionally, we tested samples of hand-operated and electric vacuum pumps. For each apparatus we recorded vacuum in millimeters of mercury using a calibrated, diaphragm-type gauge. We calculated mean maximal vacuum. Work required to produce 500 mm Hg of vacuum was determined for the hand-operated pumps. At increments of vacuum, we compared the vacuum displayed by the gauge head included with the device to our standard. RESULTS: Mean wall suction was 494 mm Hg (range, 248-655). As compared to the Mity-vac, the CMI hand-operated pump needed significantly more work to generate 500 mm Hg of vacuum. Electric pumps produced a reliable and consistent vacuum. Gauges included with the pump had an average error of 4% over 400-600 mm Hg. CONCLUSION: Wall suction was an unreliable source of vacuum. The two hand-operated pumps tested required significantly different amounts of work to achieve an adequate vacuum. Addition of a fluid trap does not affect the vacuum, and head gauges supplied with vacuum pumps are accurate.
    [Abstract] [Full Text] [Related] [New Search]