Search MEDLINE/PubMed
Title: Vacuum sources in obstetrics. Author: Wiper DW, Duchon MA, Muise KL. Journal: J Reprod Med; 1996 Jun; 41(6):444-6. PubMed ID: 8799923. Abstract: OBJECTIVE: To characterize the performance of vacuum equipment used in operative vaginal deliveries. STUDY METHOD: We measured wall suction in eight Cleveland area labor-and-delivery units. Additionally, we tested samples of hand-operated and electric vacuum pumps. For each apparatus we recorded vacuum in millimeters of mercury using a calibrated, diaphragm-type gauge. We calculated mean maximal vacuum. Work required to produce 500 mm Hg of vacuum was determined for the hand-operated pumps. At increments of vacuum, we compared the vacuum displayed by the gauge head included with the device to our standard. RESULTS: Mean wall suction was 494 mm Hg (range, 248-655). As compared to the Mity-vac, the CMI hand-operated pump needed significantly more work to generate 500 mm Hg of vacuum. Electric pumps produced a reliable and consistent vacuum. Gauges included with the pump had an average error of 4% over 400-600 mm Hg. CONCLUSION: Wall suction was an unreliable source of vacuum. The two hand-operated pumps tested required significantly different amounts of work to achieve an adequate vacuum. Addition of a fluid trap does not affect the vacuum, and head gauges supplied with vacuum pumps are accurate.[Abstract] [Full Text] [Related] [New Search]