These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Quantitation of cocaine and cocaethylene in canine serum by high-performance liquid chromatography. Author: Williams CL, Laizure SC, Parker RB, Lima JJ. Journal: J Chromatogr B Biomed Appl; 1996 Jun 07; 681(2):271-6. PubMed ID: 8811436. Abstract: A reversed-phase high-performance liquid chromatographic procedure for the determination of cocaine and cocaethylene in canine serum has been developed. The compounds were extracted from 1 ml of alkalinized canine serum with hexane. Chromatographic separation was achieved with a cyanopropyl column (250 x 4.6 mm I.D., 5 microns) using a mobile phase of acetonitrile and phosphate buffer, pH 7.40 (38:62, v/v) flowing at 1 ml/min. Eluate was monitored by a variable-wavelength UV detector set to 230 nm. The extraction procedure yields an average recovery of 99 and 96% for cocaine and cocaethylene, respectively. The between-day coefficients of variation, at 2400 ng/ml, for cocaine and cocaethylene were both 8.6% and the within-day coefficients of variation, at 400 ng/ml, for cocaine and cocaethylene were 7.3 and 8.0%, respectively. A concentration-time profile resulting from administration of 3 mg/kg cocaine and cocaethylene to the dog revealed a similar disposition between cocaine and cocaethylene, with a clearance and volume of distribution at steady-state values of 72.8 and 61.0 ml/min/kg and 2.6 and 2.7 1/kg, respectively.[Abstract] [Full Text] [Related] [New Search]