These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Inhibition of bradycardia pacing caused by far-field atrial sensing in a third-generation cardioverter defibrillator with an automatic gain feature. Author: Curwin JH, Roelke M, Ruskin JN. Journal: Pacing Clin Electrophysiol; 1996 Jan; 19(1):124-6. PubMed ID: 8848369. Abstract: The diagnostic accuracy of implantable cardioverter defibrillators may be improved by automatically adjusting gain algorithms, which in general reduce the likelihood of oversensing while maintaining the ability to detect the low amplitude signals associated with ventricular fibrillation. We present a patient with a third-generation device who developed prolonged ventricular asystole arising as a complication of the automatic gain feature. During asystole the device automatically increased sensitivity in order to prevent undersensing of ventricular fibrillation, which in this case resulted in far-field sensing of atrial activity and inhibition of ventricular pacing.[Abstract] [Full Text] [Related] [New Search]