These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: High-performance liquid chromatographic determination of finasteride in human plasma using direct injection with column switching. Author: Takano T, Hata S. Journal: J Chromatogr B Biomed Appl; 1996 Feb 09; 676(1):141-6. PubMed ID: 8852054. Abstract: A fully automated column-switching high-performance liquid chromatographic (HPLC) method was developed for the quantification of finasteride [N-(1,1-dimethylethyl)-3-oxo-4-aza-5 alpha-androst-1-ene-17 beta- -carboxamide] in human plasma. Plasma samples were diluted with an equal volume of ethylene glycol-water (40:60, v/v), then the diluted sample (150 microliters) was injected into the HPLC system without clean-up. The analyte was retained on a pretreatment column, whereas plasma proteins and other endogenous components were washed out to waste. The analyte was transferred to the analytical column in the heart-cut mode and then detected at 210 nm. A quantification limit of 1 ng/ml was attained. There was a linear relationship between peak height and drug concentration in plasma in the range 1-50 ng/ml. This method was validated and applied to the assay of plasma samples to characterize pharmacokinetic parameters in clinical studies.[Abstract] [Full Text] [Related] [New Search]