These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A novel automated assay for malate dehydrogenase isoenzymes.
    Author: Sugiuchi H, Uji Y, Shirahase Y, Okabe H.
    Journal: J Clin Lab Anal; 1996; 10(2):78-84. PubMed ID: 8852359.
    Abstract:
    We have developed a new automated method for the determination of malate dehydrogenase (MDH) isoenzymes in serum employing guanidine hydrochloride. Our proposed method showed good reproducibility; within-run precision coefficient of variations (CVs) were less than 2.5 (mean 13.6-42.9 U/L) for total MDH (T-MDH) and less than 6.7% (mean 6.3-23.5 U/L) for mitochondrial MDH (m-MDH) (n = 10). The upper detection limit of the proposed method exhibited good linearity up to 1,000 U/L for both T-MDH and m-MDH. In the proposed m-MDH reagent, the presence of up to 2,000 U/L of cytosolic MDH(c-MDH) activity had no effect on the outcome of m-MDH assay. Results of our proposed method (y) correlated well with those of the electrophoretic method (x) giving the regression equation: y = 1.46 x + 6.87 (N = 30); r = 0.99. Normal concentrations of various anticoagulants and bilirubin did not affect the assay results. Both ascorbic acid and glucose exhibited a slight positive interference with the proposed assay. Clinically, we found that m-MDH activity in serum had greater diagnostic predictive value than T-MDH activity for judging successful outcome of reperfusion therapy; the prognosis was poor when the m-MDH/T-MDH ratio was greater than 20%.
    [Abstract] [Full Text] [Related] [New Search]