These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Blood pressure monitoring during pregnancy. Accuracy of portable devices designed for obese patients.
    Author: Schwartz WJ, Rayburn WF, Turnbull GL, Christensen HD.
    Journal: J Reprod Med; 1996 Aug; 41(8):581-5. PubMed ID: 8866385.
    Abstract:
    OBJECTIVE: To compare the accuracy of three commercially available blood pressure monitoring devices having cuffs placed at different anatomic sites on obese pregnant women with large arms. STUDY DESIGN: Fifty-five obese pregnant women (body mass index > 27.3) were eligible for participation; each had an upper arm circumferences > 35 cm. The three different portable devices compared had cuffs that fit easily around either the index finger, wrist or large arm. Two recordings using each device were compared with those obtained simultaneously using a well calibrated monitor. RESULTS: Systolic, diastolic and mean arterial pressure recordings did not correlate between the monitor and devices with the cuff around the finger (r2 = .17, .17, .22), wrist (.30, .24, .33) or large arm (.44, .26, .40). The percentages of measurement differences within 5 mm Hg for the systolic, diastolic and mean arterial pressure were low for the device with the cuff around the finger (11.0%, 25.5%, 23.6%), wrist (33.0%, 46.4%, 35.5%) or large arm (38.5%, 29.4%, 46.7%). CONCLUSION: Despite their commercial appeal, none of these portable blood pressure monitoring devices was accurate for use by obese pregnant patients with large arms.
    [Abstract] [Full Text] [Related] [New Search]