These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: How to modify a volume-cycled home ventilator to satisfy a child's need for pressure-limiting and continuous positive airway pressure during spontaneous breathing. Author: Zinman R, Ness V. Journal: Respiration; 1996; 63(6):363-7. PubMed ID: 8933655. Abstract: Piston driven volume-cycled home ventilators increase work of breathing in the synchronized intermittent mandatory ventilation mode. A 2-year-old trisomy 21 patient with chronic lung disease due to recurrent aspiration pneumonia required a ventilator rate of 6-8 breaths per minute awake and 15 asleep, with peak pressure of 32 cm H2O and positive end-expiratory pressure (PEEP) of 10 cm H2O. Two circuits were designed to facilitate breathing and respiratory mechanics of his spontaneous breaths on both were compared. A spring valve on the inspiratory line pressure-limited ventilation and at the end of the expiratory line provided PEEP in both systems. The reservoir system had a 2-liter bag on the inspiratory limb of the circuit, and two one-way valves at the patient connector to direct inspiratory and expiratory flow. The continuous positive airway pressure (CPAP) system provided continuous flow with a CPAP device set to deliver a pressure 2 cmH2O higher than the PEEP valve. On the CPAP system, compared to the reservoir system, dynamic compliance was greater [1.52 (0.14 SD) ml/cm H2O/kg vs. 0.39 (0.02), p < 0.001] and resistance less [8.15 (1.26) cm H2O/l/s vs. 45.86 (0.87), p < 0.001] as measured with a PeDS machine. This is an innovative use of a CPAP device.[Abstract] [Full Text] [Related] [New Search]