These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Accuracy of home blood glucose meters during hypoglycemia. Author: Trajanoski Z, Brunner GA, Gfrerer RJ, Wach P, Pieber TR. Journal: Diabetes Care; 1996 Dec; 19(12):1412-5. PubMed ID: 8941473. Abstract: OBJECTIVE: To evaluate the accuracy of home blood glucose meters during hypoglycemia. METHODS: Six blood glucose meters-One Touch II (LifeScan, Milpitas, CA), Companion II (Medisense, Cambridge, U.K.), Reflolux (Boehringer Mannheim, Mannheim, Germany), Accutrend (Boehringer Mannheim), Elite (Bayer, Munich, Germany), and HemoCue (HemoCue, Angelholm, Sweden)-were compared with a reference method (Beckman Glucose Analyzer 2). Glucose concentrations from arterialized venous blood samples were measured using all glucose meters (whole blood) and the reference method (plasma) during hypoglycemic-hyperinsulinemic clamps in 15 subjects. RESULTS: In total, 663 blood glucose monitor readings and 119 reference values ranging from 2.28 to 3.89 mmol/l were analyzed. The correlation coefficients and the percentage of measurements within 20% and outside 40% of the reference values for each glucose meter were as follows: One Touch II: 0.91, 99.2% and 0%; Companion II: 0.81, 88.2% and 2.5%; Reflolux: 0.78, 85.0% and 0.9%; Accutrend: 0.88, 46.0% and 6.6%; Elite: 0.78, 75.6% and 4.2%; and HemoCue: 0.93, 96.6% and 0% (P < 0.001). CONCLUSIONS: There were substantial differences between the blood glucose meters during hypoglycemia, and none of the devices met the latest criteria recommended by the American Diabetes Association.[Abstract] [Full Text] [Related] [New Search]