  • Title: Improved resolution in the gel electrophoresis of proteins by a periodically interrupted electric field.
    Author: Chen N, Chrambach A.
    Journal: J Biochem Biophys Methods; 1996 Dec 30; 33(3):163-70. PubMed ID: 9029260.
    Abstract:
    The capability of the commercial gel electrophoresis apparatus with intermittent scanning of fluorescence (HPGE-1000, LabIntelligence) to provide time-dependent zone dispersion allows one to quantitate resolution. Using a model protein separation, that between phycoerythrin and fluorescein-labeled conalbumin, resolution was compared between separations conducted at a constant field strength of 80 V/cm and one conducted in 10-s pulses of the same field strength, interrupted periodically by 120 s in the absence of an electric field. Resolution was improved by a factor of two in the discontinuous application of the electric field compared to that obtained in its continuous application. Similarly, the intermittent application of 80 V/cm for 10 s, followed by 120-s pauses, gave rise to twice the resolution obtained from a continuous application of 7 V/cm.