These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Severe syncope due to the application of a pacemaker telemetry wand. Author: Preumont N, De Canniere D, Baroud I, Stoupel E. Journal: Acta Cardiol; 1996; 51(6):541-3. PubMed ID: 9124024. Abstract: We report a case of a transient defective anodal contact between pacemaker ground plate and patient's tissues (the so-called dry pocket). During unipolar-DDD pacing, mild compression of the pocket with the telemetry head caused a severe syncope, due to asystole and secondary to complete loss of ventricular capture. Normal pacemaker function resumed with bipolar pacing configuration.[Abstract] [Full Text] [Related] [New Search]