These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Evaluation of a rapid, quantitative cardiac troponin I immunoassay. Author: Kuhr LP, Baum H, Schweigert R, Hafner G, Prellwitz W, Neumeier D. Journal: Eur J Clin Chem Clin Biochem; 1997 May; 35(5):399-404. PubMed ID: 9189748. Abstract: We evaluated a rapid, quantitative immunoassay for the detection of cardiac troponin I. Coefficient of variation is between 1.29 and 13.63% for intra-assay and between 3.88 and 10.15% for inter-assay imprecision. Linearity is given up to 35 micrograms/l. Possible interfering substances (haemoglobin, bilirubin, triacylglycerol and rheuma factors) do not disturb the assay. The analyte is stable under normal storage conditions (+20 degrees C/48 h and +4 degrees C/l week) with decrease up to 30% after 3 months at -20 degrees C. Reference value for apparently healthy individuals is < 0.1 microgram/l. In plasma cardiac troponin I is measured up to 30% depressed compared to serum. Comparison with another cardiac troponin I assay (y = 0.92x + 2.42, r = 0.940) and cardiac troponin T is good with y = 6.61x - 1.94, r = 0.91 for the first generation cardiac troponin T assay and y = 5.59x - 0.68, r = 0.87 for the second generation cardiac troponin T assay. In summary, the evaluated assay is fast, easy to perform, and can be used not only in a specialized laboratory, but is also suitable for emergency laboratory or smaller laboratory units.[Abstract] [Full Text] [Related] [New Search]