These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Evaluation of Ames' "Clini-Tek". Author: Peele JD, Gadsden RH, Crews R. Journal: Clin Chem; 1977 Dec; 23(12):2238-41. PubMed ID: 923073. Abstract: Reproducibility of reading "N-Multistix" dipsticks by a semi-automated urinalysis instrument (Ames' "Clini-Tek") has been described for artifically prepared samples. Glucose, ketone, urobilinogen, and nitrite showed high reproducibility (greater than 90%) for reading multiple samples at predetermined analyte concentrations. Determination of proteinuria showed the lowest proportion of false positives (2-3%) and false negatives (0%). Determination of hemoglobinuria and bilirubinuria by dipsticks were the least reproducible. Urobilinogen showed no interference from bilirubin in concentrations up to 32 mg/liter. Precision was high for results for quality-control capsules provided by the manufacturer.[Abstract] [Full Text] [Related] [New Search]