Title: Measurements of threshold shifts observed when using normal and preformed electrodes. Author: Reuter G, Cords SM, Issing P, Lenarz T. Journal: Am J Otol; 1997 Nov; 18(6 Suppl):S26. PubMed ID: 9391585. Abstract: OBJECTIVE: To investigate the efficacy of short-term electrical stimulation with different human electrodes using high stimulation rates and different electrode designs. MATERIALS AND METHODS: Human electrodes of two different designs were implanted in cats, and electrically evoked auditory brain stem responses (EABR) were recorded. RESULTS AND DISCUSSION: Thresholds from the EABR recordings at the apical end of the electrode array were comparable in both types of implants. There were significantly higher thresholds at the basal end than at the apical end of the electrodes in both types of electrodes.[Abstract] [Full Text] [Related] [New Search]