These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Experimental study on the location of energy windows for scatter correction by the TEW method in 201Tl imaging].
    Author: Kojima A, Matsumoto M, Ohyama Y, Tomiguchi S, Kira M, Takahashi M.
    Journal: Kaku Igaku; 1997 Sep; 34(9):831-6. PubMed ID: 9394551.
    Abstract:
    To investigate validity of scatter correction by the TEW method in 201Tl imaging, we performed an experimental study using the gamma camera with the capability to perform the TEW method and a plate source with a defect. Images were acquired with the triple energy window which is recommended by the gamma camera manufacturer. The result of the energy spectrum showed that backscattered photons were included within the lower sub-energy window and main energy window, and the spectral shapes in the upper half region of the photopeak (70 keV) were not changed greatly by the source shape and the thickness of scattering materials. The scatter fraction calculated using energy spectra and, visual observation and the contrast values measured at the defect using planar images also showed that substantial primary photons were included in the upper sub-energy window. In TEW method (for scatter correction), two sub-energy windows are expected to be defined on the part of energy region in which total counts mainly consist of scattered photons. Therefore, it is necessary to investigate the use of the upper sub-energy window on scatter correction by the TEW method in 201Tl imaging.
    [Abstract] [Full Text] [Related] [New Search]