These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [In-line leukocyte depletion ov thrombocytapheresis concentrates with the Fresenius-AS-104 cell separator].
    Author: Zeiler T, Kretschmer V.
    Journal: Beitr Infusionsther Transfusionsmed; 1997; 34():110-3. PubMed ID: 9417333.
    Abstract:
    UNLABELLED: This study reports on in-line filtration of 72 platelet concentrates (PC) collected by the Fresenius AS 104 cell separator, using the new C4F sets with integrated leukocyte filters (Biofil P plus). MATERIAL AND METHODS: 72 volunteer donors, automatic counts of platelets, microscopical counting of residual leukocytes with the Nageotte chamber, GMP-140 by flow cytometrie, beta-thromboglobulin release, platelet aggregation (ADP, collagen). RESULTS AND CONCLUSION: Filtration reduced leukocytes by 98.5%. Residual leukocyte contamination remained clearly below 5 x 10(6) (mean 0.5 +/- 0.6 x 10(6), maximum 2.8 x 10(6). Platelet loss by filtration was found to be between 27.4 and 0.7% (median 8.5%). Filtration caused a significant decrease of platelet aggregability (p < 0.005), but no significant increase of beta-thromboglobulin release and only a slight decrease of GMP-140 expression. From these data can be concluded that in-line filtration was highly efficient with acceptable platelet retention. No significant platelet activation could be observed in the PC. The decrease of platelet aggregability have been due to the reduction of activated platelets which are believed to show reduced in vivo survival.
    [Abstract] [Full Text] [Related] [New Search]