These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Improved sensitivity of detection with the commercial automated gel electrophoresis (HPGE-1000) apparatus through modification of its optical system. Author: Yarmola E, Chrambach A. Journal: Electrophoresis; 1997 Oct; 18(11):1951-4. PubMed ID: 9420151. Abstract: In a representative application to a fluorescently detectable protein of commercial automated gel electrophoresis apparatus (HPGE-1000, LabIntelligence, Belmont, CA) the sensitivity of detection by fluorescence was significantly increased by elimination of the mirror below the gel tray. That increase in detection sensitivity is due to a decrease of fluorescent background noise by nearly one order of magnitude, overcompensating a decrease in signal by a factor of two. The resulting increase in signal/noise ratio, i.e., detection sensitivity, should allow for lowered sample loads by which the band width is reduced with benefits to resolution.[Abstract] [Full Text] [Related] [New Search]