Search MEDLINE/PubMed
Title: Improved high-performance liquid chromatographic assay for the determination of procainamide and its N-acetylated metabolite in plasma: application to a single-dose pharmacokinetic study. Author: Lessard E, Fortin A, Coquet A, Bélanger PM, Hamelin BA, Turgeon J. Journal: J Chromatogr Sci; 1998 Jan; 36(1):49-54. PubMed ID: 9443381. Abstract: An improved high-performance liquid chromatographic assay for the determination of procainamide and N-acetylprocainamide (NAPA) at concentrations observed up to 32 h after a single oral dose administration of procainamide to human subjects is reported. Following liquid-liquid extraction of plasma samples, procainamide, NAPA, and the internal standard (N-propionylprocainamide) are separated on a reversed-phase C8 column with retention times of 4.0, 6.7, and 13.2 min, respectively. The ultraviolet detection limit (wavelength, 280 nm) of procainamide and NAPA is 2 ng/mL (signal-to-noise ratio, 3:1), and the quantitation limit is 4 ng/mL (signal-to-noise ratio, 5:1). Intra- and interday coefficients of variation are less than 8% in the range of 20-500 ng/mL.[Abstract] [Full Text] [Related] [New Search]