Title: A comparative evaluation of the effect of pump type and heparin-coated surfaces on platelets during cardiopulmonary bypass. Author: Misoph M, Babin-Ebell J, Schwender S. Journal: Thorac Cardiovasc Surg; 1997 Dec; 45(6):302-6. PubMed ID: 9477463. Abstract: Cardiopulmonary bypass (CPB) is associated with hemostatic disorders, mainly due to platelet function defects. In the present prospective study, platelet count and GMP-140 expression on platelets were investigated to comparatively evaluate the impact of different CPB-circuit systems on platelets. The study included 61 patients undergoing coronary artery bypass grafting, randomly assigned to 3 groups. In group A a roller pump, in group B a centrifugal pump, and in group C a centrifugal pump with a Carmeda heparin-coated CPB circuit was applied. Platelet count and GMP-140 expression were analysed pre, during, and following CPB. None of the tested CPB systems did affect platelet count. The percentage of GMP-140-positive platelets increased slightly early during CPB, whereas it decreased significantly postoperatively; group differences were observed between B and C after protamin administration. An advantage of the use of centrifugal pumps and heparin-coated circuits could not be proven with the present set-up. The results suggest that the benefit of the tested systems might depend on the operative procedure and management.[Abstract] [Full Text] [Related] [New Search]