These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A comparative evaluation of the effect of pump type and heparin-coated surfaces on platelets during cardiopulmonary bypass.
    Author: Misoph M, Babin-Ebell J, Schwender S.
    Journal: Thorac Cardiovasc Surg; 1997 Dec; 45(6):302-6. PubMed ID: 9477463.
    Abstract:
    Cardiopulmonary bypass (CPB) is associated with hemostatic disorders, mainly due to platelet function defects. In the present prospective study, platelet count and GMP-140 expression on platelets were investigated to comparatively evaluate the impact of different CPB-circuit systems on platelets. The study included 61 patients undergoing coronary artery bypass grafting, randomly assigned to 3 groups. In group A a roller pump, in group B a centrifugal pump, and in group C a centrifugal pump with a Carmeda heparin-coated CPB circuit was applied. Platelet count and GMP-140 expression were analysed pre, during, and following CPB. None of the tested CPB systems did affect platelet count. The percentage of GMP-140-positive platelets increased slightly early during CPB, whereas it decreased significantly postoperatively; group differences were observed between B and C after protamin administration. An advantage of the use of centrifugal pumps and heparin-coated circuits could not be proven with the present set-up. The results suggest that the benefit of the tested systems might depend on the operative procedure and management.
    [Abstract] [Full Text] [Related] [New Search]