These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Value of positive auto controls in the gel centrifugation method]. Author: Eichler H, Kretschmer V. Journal: Beitr Infusionsther Transfusionsmed; 1994; 32():156-8. PubMed ID: 9480074. Abstract: We studied 97 samples of patients being positive in the autocontrol of the indirect antiglobulin test (IAT) in the gel system (DiaMed). In 83.2%, retesting with monospecific anti-IgG serum gave also positive results, due to a specific phenomenon caused, for example, by drug-specific antibodies (AB), warm auto-AB or allo-AB. In contrast, only 52.9% of the samples retested by the standard tube technique with polyspecific antiglobulin serum reacted positive. Only in 6 patients slightly increased cold agglutinins could be detected. None of the investigated patients showed any clinical or laboratory signs of hemolysis except one with pernicious anemia. We conclude that positive results of the autocontrol in the gel IAT should be confirmed by an additional DAT in the tube technique. If this second test shows a negative result, transfusions can take place without any restrictions.[Abstract] [Full Text] [Related] [New Search]