Search MEDLINE/PubMed
Title: Sensitive high-performance liquid chromatographic method with fluorometric detection for the simultaneous determination of gabapentin and vigabatrin in serum and urine. Author: Wad N, Krämer G. Journal: J Chromatogr B Biomed Sci Appl; 1998 Jan 23; 705(1):154-8. PubMed ID: 9498684. Abstract: Serum concentrations of the antiepileptic drug gabapentin (GBP) are usually determined by high-performance liquid chromatography (HPLC) using UV photometric detection after pre-column derivatization with 2,4,6-trinitrobenzenesulphonic acid. Vigabatrin levels in serum are determined by HPLC using fluorescence detection. Like vigabatrin (VGB), gabapentin has also a primary amine group that easily reacts with o-phthaldialdehyde reagent and produces a fluorescing substance. By the use of fluorometric detection, GBP can be determined more simply, sensitively and simultaneously with VGB. The day-to-day coefficient of variation for the determination of GBP in a pooled serum was 4.0% (n=17; serum concentration, 13.8 micromol/l) and forVGB was 3.1% (n=21; serum concentration, 26.4 micromol/l). The lower limit of detection is 0.5 micromol/l for both drugs and the method is linear up to 500 micromol/l for GBP and 1300 micromol/l for VGB.[Abstract] [Full Text] [Related] [New Search]