These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Development and validation of a sensitive method for the determination of ganciclovir in human plasma samples by reversed-phase high-performance liquid chromatography.
    Author: Campanero MA, Sadaba B, García-Quetglas E, Azanza JR.
    Journal: J Chromatogr B Biomed Sci Appl; 1998 Mar 20; 706(2):311-7. PubMed ID: 9551818.
    Abstract:
    A rapid, sensitive, specific liquid chromatographic method has been developed for the determination of therapeutic levels of ganciclovir in human plasma. Plasma (1 ml) and acyclovir (I.S.) were treated with 50% trichloroacetic acid. The supernatant was neutralized with 2 M NaOH and purified with chloroform. The aqueous phase (80 microl) was analyzed by a 3-microm Hypersil ODS C18 column with 0.04 M triethylamine-0.1 M sodium dihydrogen phosphate monohydrate as the mobile phase (1 ml/min) and ultraviolet detection at 254 nm. Calibration was linear from 50 to 10000 ng/ml. Intra- and inter-day C.V. did no exceed 6.65%. The detection limit was about 10 ng/ml.
    [Abstract] [Full Text] [Related] [New Search]