  • Title: QT dependent T wave sensing.
    Author: Böhm A, Pintér A, Préda I.
    Journal: Pacing Clin Electrophysiol; 1998 Jul; 21(7):1490-1. PubMed ID: 9670199.
    Abstract:
    A case of QT duration dependent T wave oversensing mimicking QRS undersensing and pacemaker malfunction is presented. This example indicates that interpretation of pacemaker ECG may be difficult, even in case of VVI pacemaker.