These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Cyanide determination in biological fluids using a microdiffusion method with a flow system and polarographic detection.
    Author: do Nascimento PC, Bohrer D, de Carvalho LM.
    Journal: Analyst; 1998 May; 123(5):1151-4. PubMed ID: 9709498.
    Abstract:
    The report describes a method for the automated polarographic determination of cyanide as tetracyanonickelate (II) anion complex in a gas-diffusion flow system. The volatile cyanide, existing in whole blood, plasma and urine samples, was measured after gas-diffusion using 8 x 10-5 mol l-1 hexaaminenickel solution as acceptor. The linear range of calibration, for measurements at the hanging mercury-drop electrode (HMDE), was from 0.1 to 2.0 micrograms cyanide with r = 0.998. The RSD was, respectively, 3.4 and 1.2% (n = 5) for 0.4 microgram cyanide measured with and without the flow-system configuration. Detection limits of 7.4 microgram l-1 were calculated using the flow system and the method was compared with the classical method using Cavet flasks. Parameters that affect the cyanide determination in the proposed method, such as acceptor solution, pH, flow rate and temperature, were investigated.
    [Abstract] [Full Text] [Related] [New Search]