These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: On-line computation of cardiac output with the thermodilution method, using a digital minicomputer. Author: Snoeckx LH, Verheyen JL, Van De Water A, Lewi P, Reneman RS. Journal: Cardiovasc Res; 1976 Sep; 10(5):556-64. PubMed ID: 971471. Abstract: In the present study, several techniques for calculating cardiac output were evaluated in order to find an accurate technique suitable for on-line digital computation. The thermodilution method was evaluated by different calculation techniques and by comparing these techniques with electromagnetic flowmeter values. Comparison of thermal dilution curves, manually calculated by the log-normal and the exponential assumption, showed a correlation coefficient of 0.978 between the two methods, the former values being 11% higher. The exponential method correlated very well with the technique, using a correction factor for injection errors (r = 0.999). Since the exponential technique correlated well with the electromagnetic values (r = 0.84) and since it was less complicated than the technique, using correction factors, this technique was chosen for automation. Comparison of exponential calculated thermodilution values with a digital computer and manually, showed a correlation coefficient of 0.991. Therefore, it was concluded that on-line computation of thermodilution curves improves the applicability of the termodilution techniques as a means of measuring cardiac output.[Abstract] [Full Text] [Related] [New Search]