These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Long-term seal provided by some root-end filling materials. Author: Wu MK, Kontakiotis EG, Wesselink PR. Journal: J Endod; 1998 Aug; 24(8):557-60. PubMed ID: 9759020. Abstract: A tight and long-lasting seal of root-end fillings is of prime clinical importance. A hundred standard bovine root sections, each 3 mm high and with a central pulp lumen of 2.6 mm in diameter, were filled with five commonly used or potential root-end filling materials. At 24 h, or at 3, 6, or 12 months after filling, leakage along these filling materials was determined under a low headspace pressure of 10 kPa (0.1 atm) using a fluid transport model. During the first 3 months, the percentage of gross leakage (> 20 microliters day-1) increased noticeably for Tytin amalgam (from 20 to 100%) and Super-EBA (from 0 to 55%), whereas it decreased noticeably for mineral trioxide aggregate (MTA; from 55% to 0%). Thereafter, the increased leakage of amalgam and Super-EBA decreased with time, whereas the improved seal of MTA was maintained until the end of the experiment. At 3-, 6-, and 12-month time intervals, both glass ionomer cements (Fuji II and Hi Dense) and MTA showed less leakage than the conventional amalgam and Super-Eba, of which amalgam leaked more.[Abstract] [Full Text] [Related] [New Search]