Search MEDLINE/PubMed
Title: Pulse oximetry in severe carbon monoxide poisoning. Author: Hampson NB. Journal: Chest; 1998 Oct; 114(4):1036-41. PubMed ID: 9792574. Abstract: STUDY OBJECTIVES: To evaluate the accuracy and quantitate the error of pulse oximetry measurements of arterial oxygenation in patients with severe carbon monoxide (CO) poisoning. DESIGN: Retrospective review of patient clinical records. SETTING: Regional referral center for hyperbaric oxygen therapy. PATIENTS: Thirty patients referred for treatment of acute severe CO poisoning who demonstrated carboxyhemoglobin (COHb) levels >25%, with simultaneous determinations of arterial hemoglobin oxygen saturation by pulse oximetry (SpO2) and arterial blood gas (ABG) techniques. MEASUREMENTS AND RESULTS: COHb levels and measurements of arterial oxygenation from pulse oximetry, ABG analysis, and laboratory CO oximetry were compared. SpO2 did not correlate with COHb levels. SpO2 consistently overestimated the fractional arterial oxygen saturation. The difference between arterial hemoglobin oxygen saturation (SaO2) calculated from ABG analysis and SpO2 increased with increasing COHb level. CONCLUSIONS: Presently available pulse oximeters overestimate arterial oxygenation in patients with severe CO poisoning. An elevated COHb level falsely elevates the SaO2 measurements from pulse oximetry, usually by an amount less than the COHb level, confirming a prior observation in an animal model. Accurate assessment of arterial oxygen content in patients with CO poisoning can currently be performed only by analysis of arterial blood with a laboratory CO-oximetry.[Abstract] [Full Text] [Related] [New Search]