These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Fast and thermal neutron profiles for a 25-MV x-ray beam.
    Author: Price KW, Nath R, Holeman GR.
    Journal: Med Phys; 1978; 5(4):285-9. PubMed ID: 98695.
    Abstract:
    High-energy x-ray radiotherapy machines generate neutrons by photonuclear reactions in the target and the treatment head and expose the patient to a neutron flux. In order to evaluate the neutron exposure quantitatively, fast and thermal neutron profiles for 25-MV x-ray beams of the Sagittaire accelerator have been measured. An activation technique, using the reactions 31P(n, gamma)32P (thermal neutrons) and 31P(n, p)31Si (fast neutrons, E greater than 0.7 MeV), has been developed to measure fast- and thermal-neutron fluxes in an intense high-energy photon flux. The sensitivity of this activation detector to high-energy photons, which has plagued many previous neutron measurements, was carefully measured and found to be less than 4%. Neutron fluxes for various photon field sizes ranging from 5 X 5 cm to 30 X 30 cm have been measured. The fast-neutron profiles were observed to have rounded edges and the thermal fluxes were found to be relatively uniform. In the central part of the x-ray beam, the ratio of neutron dose equivalent to photon absorbed dose was found to be between 0.2% and 0.5%. Outside of the photon field, the ratio of neutron dose equivalent to the central-axis photon absorbed dose was 0.12%.
    [Abstract] [Full Text] [Related] [New Search]