These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Automatic motion correction in dynamic radionuclide renography using image registration.
    Author: Lee KJ, Barber DC.
    Journal: Nucl Med Commun; 1998 Dec; 19(12):1159-67. PubMed ID: 9885806.
    Abstract:
    Time-activity curves from dynamic renograms can be analysed to yield important quantitative parameters. However, accurate results are dependent on curves being free from artefacts caused by patient movement. We describe a method to correct for both translational and rotational motion during renography. The method is image-based and does not require markers. Sequential dynamic frames are registered using an affine transform from which rotational and translational components are extracted by singular-value decomposition. Computer simulations demonstrate correction to < 1 pixel and < 1 degree. The method is also fast, with renograms comprising 75 frames of 64 x 64 pixels taking under 1 min for correction on a Sun SPARC 5.
    [Abstract] [Full Text] [Related] [New Search]